MAXX TP5 C: Portable sampler as compact device for fully automatic sampling according to the vacuum principle. Battery-operated 12V/10Ah.
Features/ Specifications:
• Type: Portable sampler
• Housing: PE/ PC (GF10)
• Thermostatic control: Insulated lower part (sample compartment, insulation thickness 40 mm); Option: freezer packs (200x10x8 mm); Option: compressor cooling (12V/115V/230V)
• Control: Microprocessor control, Sleep-Mode (<5mA), power supply 8-16 V; foil keyboard (with keys 0-9, ESC, ENT, cursor), graphical display (128*64 Pixel), back lit
• Data logger: 3000 entries, non-volatile data memory; storage of sampling and malfunction data like sample extractions, bottle changes, messages, external signals. Optional with WEB-board 100 MB (2 Year ring memory-FIFO at 1 min interval)
• Programming: 12 freely programmable user programs, with function to link programs.
• Program start options: IMMEDIATELY; DATE/TIME; WEEKDAY/TIME; BY AN EXTERNAL SIGNAL
• Program End/Stop options: End of sampling program; AFTER 1 RUN; AFTER X RUNS; CONTINUOUS OPERATION; DATE/TIME
• Pause mode: Interruption of program run at any time
• Overfilling protection: Adjustable from 1 – 999 samples/bottles
• Interval setting: 1 min. to 99 h 59 min. in steps of 1 minute
• Pulse setting: 1 to 9999 pulses/sample
• Manual sample extraction: Possible at any time without interrupting the current program run
• Program protection: Up to 5 years after voltage loss
• Interface: Mini-USB, RS422, RS 232; Optional: Ethernet RJ45
• Communication: LAN/ WLAN/ GPRS-UMTS:
optional:
1. Connection via USB and PC:
• maxxwareConnect® has to be installed on the PC
• Connection to the sampler via USB/MiniUSB cable
• remote control of the sampler
• visualization of downloaded data
• download and saving of data as PDF, CSV or XLSX format
• print-out of data directly as PDF format
• backup of all preprogramed programs from the sampler
• setting and saving of programs in offline mode. Upload in online mode
• Read out, changing, saving or upload of all sampler programs (1-12)
• recovery of saved programs.
or alternatively:
2. LAN Modul RJ45 via TCP/IP and IE-Browser:
• ARM9-SoC
• 32MB RAM
• 100 MB Data Memory (2 Year ring memory-FIFO at 1 min interval)
• Linux OS
• TCP/IP (RJ45)
• recording of all CPU Data (like data of sampling cycle, bottle report, error log, temperature etc.)
• visualization via Web interface
• Data-export (PDF, CSV, XLS)
• E-Mail error messaging
or alternatively:
3. LAN Modul RJ45 + GPRS/UMTS Router:
• ARM9-SoC
• 32MB RAM
• 100 MB Data Memory (2 Year ring memory-FIFO at 1 min interval)
• Linux OS
• TCP/IP (RJ45)
• recording of all CPU Data (like data of sampling cycle, bottle report, error log, temperature etc.)
• visualization via Web interface
• Data-export (PDF, CSV, XLS)
• E-Mail error messaging additionaly + fully integrated router (industrial standard) + UMTS/GPRS + SIM card holder + E-Mail error messaging + antenna
• Languages: Multi-language, selectable
• Signal inputs: 2x analogue: 0/4-20 mA; 8x digital (flow, event, 1 input can be programmed freely); Option: expandable with 4x digital, 3 inputs can be programmed freely, and 8x analogue 0-20 mA or 0-10 V, impulslength 60ms, switching level 7-24 V,
max. working restistance 500 Ohm, max. length of signalcable 30 m.
• Signal outputs/ status messages: 8 digital outputs, 1x of them as collective malfunction message (Relay optional); Option: expandable with 8 digital, 5 are freely programmable (in total 6 messages).
• Sampling method: Vacuum system 20-350 ml (Standard); Option: Vacuum VAR (variable) system 5-250 ml; Option: Peristaltic pump
• Single sample volume accuracy: Vacuum system: < 2,5% or +- 3 ml; peristaltic pump: +- 5% or +- 5 ml
• Suction height: Max. 6,5 m (at 1013h Pa); Option: 8 m
• Pumping speed: >0,5 m/s at suction height up to 5 m (at 1013h Pa); pump capacity can be adjusted electronically
• Suction hose: PVC, L=5 m, ID=10 mm
• max. hose length: 30 m
• Sampling modes: time-related, flow-dependent, event-related and manual sample extraction. Option: flow-proportional (for Peristaltic Pump as standard)
• Bottle variants: 1 x 10 L PE/ 1 x 25 L PE/ 2 x 13 L PE/ 4 x 5 L PE/ 16 x 1 L PE (incl. freezer packs)/ 24 x 1 L PE (standard version)
• Overall dimensions (Hxwxd): 787 x 510 x 468 mm (Insulating box passive) / 1028 x 550 x 468 mm (Insulating box active, with compressor cooling)
• Weight Approx.:
- 25 kg (24x1 L, Isobox with passive cooling)
- Approx. 40 kg (24x1 L, Isobox with compressor cooling, device incl. battery, empty bottles but no suction hose)
• Power supply: Sampler: 12 V/10Ah lead storage battery (maintenance-free, leak proof); 115V or 230V operation by means of battery charger in buffer mode. Range 11-14V; power consumption max. 30 W; Cooling Box: 230V 50Hz (115V by request) or 12V battery (solar battery with at least 90 Ah).
• Power requirement/ number of samples: Sampler: Up to 2000 sample extractions per battery charge, according to ambient conditions. Cooling Box: Power requirement with option "active cooling" approx. ...according to ambient conditions aprox. 50W. (with 90 Ah battery, 20°C ambient, sampling 3x/h = running time of cooling system aprox. 49h).
• Ambient temperature: 0 – 45°C
• Sample temperature: 0 – 40°C
• Standards: CE, Sampling according to ISO 5667-10, EN 16479
• Wetted materials: PC, PVC, Silicone, PS, PE