Wall-mounted device which can be combined with any composite container or refrigerator. With capacitive level sensor (non-contact) for industrial applications.
Features:
• New, direct connection via USB of an external multiparameter portable meter pH/Cond/Temp (Option).
• Flexible application and combination possibilities.
• Highly accurate sample volume.
• Clear operating structure and simple programming.
• LAN/WLAN/GPRS communication (Option).
• Easy cleaning.
• Modern and ergonomic design.
Specifications:
• Type Wall-mounted sampler.
• Housing PS / PC (GF10).
• Thermostatic control: No.
Option: refrigerator.
• Control: Microprocessor control, Sleep-Mode (<5mA), power supply 8-16 V; foil keyboard (with keys 0-9, ESC, ENT, cursor), graphical display (128*64 Pixel), back lit.
• Data logger: 3000 entries, non-volatile data memory; storage of sampling and malfunction data like sample extractions, bottle changes, messages, external signals. Optional with WEB-board 100 MB (2 Year ring memory-FIFO at 1 min interval).
• Programming 12 freely programmable user programs, with function to link programs.
• Program start options: IMMEDIATELY; DATE/TIME; WEEKDAY/TIME; BY AN EXTERNAL SIGNAL.
• Program End/Stop options: AFTER 1 RUN; AFTER X RUNS; CONTINUOUS OPERATION; DATE/TIME.
• Pause mode: Interruption of program run at any time.
• Overfilling protection: Adjustable from 1–999 samples/bottles.
• Interval setting: 1 min. to 99 h 59 min. in steps of 1 minute.
• Pulse setting: 1 to 9999 pulses/sample.
• Manual sample extraction: Possible at any time without interrupting the current program run.
• Program protection: Up to 5 years after voltage loss.
• Interface: Mini-USB, RS 232; Optional: Ethernet RJ45, SDI-12.
• Communication LAN / WLAN / GPRS-UMTS:
optional:
1. Connection via USB and PC:
• maxxwareConnect® has to be installed on the PC.
• Connection to the sampler via USB/MiniUSB cable.
• remote control of the sampler
• visualization of downloaded data
• download and saving of data as PDF, CSV or XLSX Format
• print-out of data directly as PDF Format
• backup of all preprogramed programs from the sampler
• setting and saving of programs in offline mode. Upload in online mode
• Read out, changing, saving or upload of all sampler programs (1-12 )
• recovery of saved programs.
or alternatively:
2. LAN Modul RJ45 via TCP/IP and IE-Browser:
• ARM9-SoC
• 32MB RAM
• 100 MB Data Memory ((2 Year ring memory-FIFO at 1 min interval)
• Linux OS
• TCP/IP (RJ45)
• recording of all CPU Data (like data of sampling cycle, bottle report, error log, temperature etc.)
• visualization via Web interface
• Data-export (PDF, CSV, XLS)
• E-Mail error messaging
or alternatively:
3. LAN Modul RJ45 + GPRS/UMTS Router:
• ARM9-SoC
• 32MB RAM
• 100 MB Data Memory ((2 Year ring memory-FIFO at 1 min interval)
• Linux OS
• TCP/IP (RJ45)
• recording of all CPU Data (like data of sampling cycle, bottle report, error log, temperature etc.)
• visualization via Web interface
• Data-export (PDF, CSV, XLS)
• E-Mail error messaging additionaly
+ Fully integrated Router (industrial standard)
+ UMTS / GPRS
+ SIM card holder
+ E-Mail error messaging
+ antenna.
• Languages Multi-language, selectable
• Signal inputs: 2 x analogue: 0/4-20 mA, 8 x digital (flow, event, 1 inputs can be programmed freely); Option: expandable with 4x digital, 3 inputs can be programmed freely, and 8x analogue 0- 20 mA or 0-10 V, Impulslength 60ms, switching level 7-24 V, max. working restistance 500 Ohm, max. length of signalcable 30 m.
• Signal outputs / status messages: 8 digital outputs, 1x of them as collective malfunction message (Relay optional); Option: expandable with 8 digital, 5 are freely programmable (in total 6 messages).
• Sampling method: Vacuum-System 20-350 ml.
• Single sample volume accuracy: Vacuum system: < 2,5 % or +- 3 ml.
• Suction height: Max. 6,5 m (at 1013hPa) optional 8,5 m or 15 m (Power Booster).
• Pumping speed: >0,5 m/s at suction height up to 5 m (at 1013h Pa); pump capacity can be adjusted electronically.
• Suction hose: PVC, L=5 m, ID=10 mm. max. hose length 30 m.
• Sampling modes: Time-related, flow-dependent, event- related and manual sample extraction.
• Bottle variants: Composite container.
• Overall dimensions Sampler (hxwxd): 362 x 442 x 222 mm.
• Weight: Approx. 10 kg.
• Power supply: 230 V / 115 V /AC.
• Power requirement: Approx. 25 VA.
• Ambient temperature: 0 – 45° C.
• Sample temperature: 0 – 40° C.
• Standards: CE Sampling according to ISO 5667-10, EN16479.
• Wetted materials: PC, PVC, Silicone, PS, PE.
• Accessories: Capacitive level sensor (non-contact) for industrial applications.
Delivery Contents:
Wall-mounted device which can be combined with any composite container or refrigerator. With capacitive level sensor (non-contact) for industrial applications, without thermostatic.