MN NANOCOLOR® VARIO C2 M, Programmable heating block with large holes
New
Description
Programmable heating block for chemical-analytical digestions with 8 holes for test tubes with Ø16 mm OD (outer diameter) and 2 holes for reaction tubes Ø 22 mm.
Features:
• High-speed COD, total nitrogen and total phosphorus in 30 minutes
• Preprogrammed standard programmes for all routine digestions
• Easy to programme for user-defined methods
• External temperature control and calibration with the NANOCOLOR® T-Set
• Simultaneous digestion of 24 samples in 2 independently usable heating units
• 5 preprogrammed temperatures 70 / 100 / 120 / 148 / 160 °C
• 4 preprogrammed heating times ½ h / 1 h / 2 h / cont.
• 7 free memory locations for individual temperature settings from30 – 160 °C (1 °C increments).
• 8 free memory locations for individual time settings from 0:01 – 99:59 h (0:01 h increments)
• Very short warm-up times (from 20 to 100 °C within 5 minutes).
• High-speed COD, total nitrogen, total phosphorus and total metals in only 30 minutes.
• Extra thick safety covers on the surface of the heating blocks as protection against contact (10 mm, exchangeable).
• Stable integrated protective coverings.
• High temperature stability.
• Electronical protection against excess temperature.
Specifications:
• Type: Programmable heating blocks for chemical-analytical digestions with 8 holes for test tubes with Ø16 mm OD (outerdiameter) and 2 holes for reaction tubes Ø 22 mm; integrated protective lid.
• Display: Coloured, backlit LCD touchscreen.
• Operation: Menu guidance via touchscreen.
• Temperatures: 5 preprogrammed temperatures (70 / 100 / 120 / 148 / 160 °C); 7 free memory locations for individual temperature settings.
• Temperature range: 30 – 160 °C (1 °C increments).
• Temperature stability: ± 1 °C (according to DIN, EN, ISO and EPA methods).
• Warm-up times: from 20 °C to 160 °C within 10 minutes
• Heating times: 4 preprogrammed heating times (½ h / 1 h / 2 h / cont.); 8 free memory locations for individual time settings.
• Time range: 1 min – 999 mi (1 min increments).
• Safety: Replacable safety cover as contact protection; Lockable protective lid; Overheating protection.
• Interface: bidirectional serial RS 232, USB A (master) and USB B (slave).
• Analytical quality control: With NANOCOLOR® T-Set (optionally fully automatic calibration and preparation of test certificates for instrument control and monitoring according to ISO 9001).
• Update: Via USB/RS232 and PC.
• Power supply: 110–230 V~, 50/60 Hz.
• Power consumption: 250/500 VA.
• Dimensions: 169 x 282 x 146 mm.
• Weight: approx. 2.0 kg.
• Marking: CE.
• Warranty: 2 years.
Price reductions